E-Catalog > เครื่องมือวัด ทดสอบ และควบคุม > Air Differentail Pressure Transmitter ( Type 699 )
PAGE VIEW : 4,064
Product Information :
Name :
Air Differentail Pressure Transmitter ( Type 699 ) Category :
- การวัดและทดสอบ, เครื่องมือวัดและทดสอบ
Sub - Category :
- เครื่องมือวัด ทดสอบ และควบคุม
Brand :
Huba Model :
( Type 699 )Product Type :
- สินค้า, ผลิตภัณฑ์
Short Description :
Relative, and differential pressure
transmitter type 699
Pressure range
-1 ... 1 mbar / 0 ... 0.3 - 50 mbar
Product Detail :
Technical overview | ||||||
Pressure range | ||||||
Relative and differential -1 ... 1 mbar / 0 ... 0.3 - 50 mbar | ||||||
Operating conditions | ||||||
Medium Air and neutral gases | ||||||
Medium / ambient 0 ... +70 ÀC | ||||||
Temperature | ||||||
Tolerable overload on one side | ||||||
Application at over pressure range | ||||||
Application at under pressure range | ||||||
Storage -10 ... +70 ÀC | ||||||
No condensation | ||||||
> 3 mbar P1 = -4 mbar P2 = -100 mbar | ||||||
< 3 mbar | P1 = 50 mbar | P2 = | 4 mbar | |||
> 3 mbar | P1 = 100 mbar | P2 = | 4 mbar | |||
< 3 mbar | P1 = -4 mbar | P2 = | -50 mbar | |||
Rupture pressure ambient temperature 2 x overload 70 ÀC 1.5 x overload | ||||||
Materials in contact with medium | ||||||
Sensor Ceramic Al2O3 (96%) | ||||||
Diaphragm Silicone | ||||||
Housing Polycarbonat PC | ||||||
Electrical overview | ||||||
Output 1) Power supply 1) Load Current consumption 2) | ||||||
2 wire 4 ... 20 mA 8.0 ... 33 VDC < supply voltage - 8 V [Ohm] < 20 mA | ||||||
0 ... 10 V 13.5 ... 33 VDC / 24 VAC μ15% > 10 kOhm < 10 mA | ||||||
0 ... 20 mA 13.5 ... 33 VDC / 24 VAC μ15% < 500 Ohm < 30 mA | ||||||
3 wire | ||||||
4 ... 20 mA 13.5 ... 33 VDC / 24 VAC μ15% < 500 Ohm < 30 mA | ||||||
0 ... 5 V 3) 6.5 ... 33 VDC / 24 VAC μ15% > 10 kOhm < 10 mA | ||||||
Filter Response time switchable by off / 0.2s / 1s / 5s / 20s | ||||||
Polarity reversal protection Short circuit proof and protected against polarity reversal. Each connection is protected against crossover up to max. supply voltage. | ||||||
Dynamic response | ||||||
Response time < 20 ms | ||||||
Load cycle < 10 Hz | ||||||
Protection standard | ||||||
Without cover IP 00 | ||||||
With cover IP 54 | ||||||
IP 65 | ||||||
Display | ||||||
LCD Display Double spaced per 8 digit alphanumeric At additional backlight LCD-Display 30 mA current consumption | ||||||
Module MODBUS | RTU RS-485 | |||||
Ranges of adjustment | ||||||
The zero piont is adjustable by reset button. | ||||||
The Full scale is adjustable by DIP-Switch and additional by the turbopoti. | ||||||
Adjustability | ||||||
Optional version with self configurable parameters (see order code selection table) | ||||||
Electrical connection | ||||||
Screw terminals for wire and stranded conductors up to 1.5 mm2 | ||||||
Cable gland with built-in strain relief PG11 | ||||||
Pressure connection | ||||||
Connection pipe | ± 6.2 mm | |||||
Mounting instructions | ||||||
Installation arrangement | Recommendation: | Vertical, with pressure connections downwards | ||||
Mounting | Mounting bracket (integrated in case) | |||||
Tests / Admissions | ||||||
UL | ||||||
Electromagnetic compatibility | CE-conformity acc. EN 61326-2-3 | |||||
Weight | ||||||
Without display | ~ 90 g | |||||
With display | ~ 100 g | |||||
Packaging | ||||||
Single packaging in cardboard | ||||||
Multiple packaging | 20 / 40 / 120 |
1 2 3 4 5 6 7 8 9 10 | |||||||||||
Order code selection table 699. X X X X X X X X X X | |||||||||||
Pre-adjustment | max. range | 9 | |||||||||
middle range | B | ||||||||||
min. range | C | ||||||||||
Signal range selectable | Pressure range of 0 % bis 100% fs | 1 | |||||||||
Pressure range of -10 % bis 90% fs | 2 | ||||||||||
Pressure range of -100 % bis 100% fs | 3 | ||||||||||
Pressure range selectable | mbar (hPa) Pa mmWS inH O 2 0 ...0.3/0.5 30/50 3/5 0.1/0.2 1) |
0 | |||||||||
0 ... 0.3/0.5/1 30/50/100 3/5/10 0.1/0.2/0.3 2) | 1 | ||||||||||
0 ... 0.5/1/3 50/100/300 5/10/30 0.3/0.5/1 1) | 2 | ||||||||||
0 ... 1/3/5 100/300/500 10/30/50 0.5/1/2 1) | 3 | ||||||||||
0 ... 3/5/10 300/500/1000 30/50/100 1/2/3 1) | 4 | ||||||||||
0 ... 5/10/16 500/1000/1600 50/100/160 2/3/5 1) | 5 | ||||||||||
0 ... 10/16/25 1000/1600/2500 100/160/250 3/5/10 1) | 6 | ||||||||||
0 ... 16/25/50 1600/2500/5000 160/250/500 5/10/20 1) | 7 | ||||||||||
Pressure unit | mbar | 0 | |||||||||
hPa | 4 | ||||||||||
Pa | 2 | ||||||||||
kPa | 5 | ||||||||||
mmWS | 3 | ||||||||||
inH O | 6 | ||||||||||
Output signal / adjustment | Linear without Filter dual DIP-Switch | 1 | 0 | ||||||||
Linear with Filter (transposable) tenfold DIP-Switch | 2 | ||||||||||
Square root extracted without Filter dual DIP-Switch | 4 | 0 | |||||||||
Square root extracted with Filter (transposable) tenfold DIP-Switch | 3 | ||||||||||
Output / power supply | 0 ... 10 V 13.5 ... 33 VDC / 24 VAC μ 15 % (3 wire) | 1 | |||||||||
0 ... 20 mA 13.5 ... 33 VDC / 24 VAC μ 15 % (3 wire) | 3 | ||||||||||
4 ... 20 mA 13.5 ... 33 VDC / 24 VAC μ 15 % (3 wire) | 4 | ||||||||||
4 ... 20 mA 8.0 ... 33 VDC (2 wire) | 5 | ||||||||||
Output signal complimentary selectabel, at delivery no pre-adjustment | 9 | 2 | 6 | ||||||||
Option | without display | 0 | |||||||||
with display in pressure unit chosen above | 1 | ||||||||||
with display in % fs | 2 | ||||||||||
with module MODBUS | 23 | 1 | 3 | ||||||||
Pressure connection / Pressure orifices | Connection pipe ± 6.2 mm without pressure orifice | 1 | |||||||||
Connection pipe ± 6.2 mm pressure orifice on P1 | 2 | ||||||||||
Connection pipe ± 6.2 mm pressure orifice on P2 | 3 | ||||||||||
Connection pipe ± 6.2 mm pressure orifice on P1 and P2 | 4 | ||||||||||
Accessories / Connection Kit | IP 54 without | 0 | |||||||||
IP 54 with connection kit (metal), 90À angled including tube 2 m long (Fig. 1) | 1 | ||||||||||
IP 54 with connection kit (plastic), straight including tube 2 m long (Fig. 2) | 2 | ||||||||||
IP 65 without | 3 | ||||||||||
IP 65 with connection kit (metal), 90À angled including tube 2 m long (Fig. 1) | 4 | ||||||||||
IP 65 with connection kit (plastic), straight including tube 2 m long (Fig. 2) | 5 | ||||||||||
Pressure range variation Indicate W and state range on order (e.g.: W0 ... + 8mbar/OUT1...6V) | 9 | 1 | W |